WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 1... — FDA record Z-1102-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 4, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device Type Part Number XXS 10-702-080095-0 XS 10-702-095115-0 Short 10-702-115150-0 Medium 10-702-140200 Long 10-702-190300-0
Reason
Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.
Classification
Class II
Recalling firm
Wishbone Medical, Inc.
Quantity
195
Agency-reported status
Ongoing
Source record identifier
Z-1102-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityXXS, Part No. 10-702-080095-0 UDI-DI B5621070208009500 Lot 1048176, 1048221 XS, Part No. 10-702-095115-0 UDI-DI B5621070209511500 Lot 1048177 Short, Part No. 10-702-115150-0 UDI-DI B5621070211515000 Lot 1048178, 1048222, 1048223 Medium, Part No. 10-702-140200 UDI-DI B562107021402000 Lot 1048179 Long, Part No. 10-702-190300-0 UDI-DI B5621070219030000 Lot 1048180

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of South Carolina.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 15, 2023
Recall initiation date
Jan 4, 2023
Center classification date
Feb 7, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records