RipCord Syndesmosis Button, REF: STA001K — FDA record Z-1101-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 22, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
RipCord Syndesmosis Button, REF: STA001K
Reason
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.
Classification
Class II
Recalling firm
TriMed Inc.
Quantity
76
Agency-reported status
Ongoing
Source record identifier
Z-1101-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00842188123849, Lot: 85305-01

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of IL, PA, TX, GA, TN, AL, MO, FL, AR, LA, HI, OH, TX, KS

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 15, 2023
Recall initiation date
Dec 22, 2022
Center classification date
Feb 7, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.