The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered ... — FDA record Z-1100-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 8, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered components intended for imaging and analysis of soft-tissue and vasculature using combined optoacoustic (photoacoustic) and ultrasound (US) imaging.
Reason
The MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).
Classification
Class II
Recalling firm
Ithera Medical Gmbh
Quantity
23 units (6 units US)
Agency-reported status
Ongoing
Source record identifier
Z-1100-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel numbers MSOT Acuity Echo (Research System) MSOT Acuity Echo CE MSOT Acuity Echo CE2 UDI's: (01)04262380070004 (01)04262380070004(21)22102 (01)04262380070004(21)22203 (01)04262380070004(21)21901 (01)04262380070004(21)22101 (01)04262380070004(21)22104 (01)04262380070004(21)22105 (01)04262380070004(21)22106 (01)04262380070004(21)22107 (01)04262380070004(21)22201 (01)04262380070004(21)22202 (01)04262380070004(21)22204 (01)04262380070004(21)22205 (01)04262380070004(21)22206 (01)04262380070004(21)22301 (01)04262380070004(21)22401

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Germany, Italy, Netherlands, Portugal, Switzerland, and United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 19, 2025
Recall initiation date
Jan 8, 2025
Center classification date
Feb 7, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.