Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U120600... — FDA record Z-1098-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 4, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090
Reason
Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.
Classification
Class II
Recalling firm
Medtronic Inc
Quantity
442
Agency-reported status
Ongoing
Source record identifier
Z-1098-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI (GTIN): 00763000547332 00763000547271 00763000547318 00763000547264 00643169796225 00643169796300 00643169796294 00643169796232 00643169796287 00763000547325 00643169796300 00763000547257 Serial/Lot numbers C112511 C113253 C113795 C117416 C117482 C121084 C123953 C133327 C133329 C134478 C138060 C150325

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 21, 2026
Recall initiation date
Dec 4, 2025
Center classification date
Jan 15, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records