Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system — FDA record Z-1092-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 5, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
Reason
Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
Classification
Class II
Recalling firm
Luminex Corporation
Quantity
24 units
Agency-reported status
Ongoing
Source record identifier
Z-1092-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers for 10-0000-07: 17340003 19044007 20133004 18128002 18214004 12061048 18191010 13193254 13051068 21029902 16277059 19233005 11250096 14069051 18255003 *18039010 *20212006 *19317004 *19206009 *18057008 Serial Numbers added 6.23.22 for 10-0000-07R: 18191010 13051068 14069051 *12198198 *12278344 *14322267 *12271338 Serial Numbers added 7.14.22 for 10-0000-07R: 18226007 and 16242043 Serial Numbers added 7.14.22 for 10-0000-07: 17062030, 18235008 and 19192006

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 25, 2022
Recall initiation date
Apr 5, 2022
Center classification date
May 13, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records