ATLAS Stim Headbox-NK For use only with the ATLAS Neurophysiology System. CAUTION--INVESTIGATIONAL DEVICE Limited by ... — FDA record Z-1091-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Mar 14, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ATLAS Stim Headbox-NK For use only with the ATLAS Neurophysiology System. CAUTION--INVESTIGATIONAL DEVICE Limited by Federal (or United States) law to investigational use. Only for use in Non-Significant Risk studies under the supervision of an IRB.
Reason
Atlas Stim Headbox (ASHB) was delivered to US customers under a premise that the device had an Investigational Device Exemption (IDE) and was for research use only, but the manufacturer does not have an approved IDE by the FDA. The product has not undergone electrical safety testing and has not received regulatory clearance.
Classification
Class II
Recalling firm
Neuralynx Inc
Quantity
15
Agency-reported status
Ongoing
Source record identifier
Z-1091-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityASHB Part No. 31-0601-0077 (SN0001, SN0002, SN0004, SN0005, SN0006, SN0007, SN0008, SN0009, SN0010, SN0012); ASHB Part No. 31-0601-0132(SN0003, SN0005, SN0006); ASHB Part No. 31-0601-0089(SN0004, SN0006);

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S. Nationwide distribution in the states of AL, CA, IL, MN, NC, RI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 25, 2022
Recall initiation date
Mar 14, 2022
Center classification date
May 13, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records