Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS... — FDA record Z-1089-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 16, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8) LEX LAP HYST, REF DYNJ902030J; 9) LITHOTOMY-SLINGS-LF, REF DYNJ905274D; 10) GYN LAPAROSCOPY, REF DYNJ905485J; 11) ROBOTIC HYSTERECTOMY, REF DYNJ906380C; 12) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ910634; 13) LAVH GYN/ONC, REF DYNJ910927.
Reason
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Classification
Class II
Recalling firm
Medline Industries, LP
Quantity
1457 units
Agency-reported status
Ongoing
Source record identifier
Z-1089-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF CDS980754T: UDI/DI 10198459121814 (EA) 40198459121815 (CS), Lot Number 25EMD795; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBR237; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBP591; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25DBM353; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25CBL184; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25BBL633; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25FBK918; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25CBJ496; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25FBF371; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25EBK493; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25DBH895; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25DBC272; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBU642; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBJ298; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25BBU206; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25FBA261; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25CBN230; REF DYNJ901829T: UDI/DI 10198459231384 (EA) 40198459231385 (CS), Lot Number 25FDB398; REF DYNJ901829T: UDI/DI 10198459231384 (EA) 40198459231385 (CS), Lot Number 25DDA191; REF DYNJ901829T: UDI/DI 10198459231384 (EA) 40198459231385 (CS), Lot Number 25BDB095; REF DYNJ902030J: UDI/DI 10195327391249 (EA) 40195327391240 (CS), Lot Number 25DBJ371; REF DYNJ905274D: UDI/DI 10195327393014 (EA) 40195327393015 (CS), Lot Number 25BBS130; REF DYNJ905485J: UDI/DI 10198459202025 (EA) 40198459202026 (CS), Lot Number 25FBJ689; REF DYNJ905485J: UDI/DI 10198459202025 (EA) 40198459202026 (CS), Lot Number 25EBP700; REF DYNJ905485J: UDI/DI 10198459202025 (EA) 40198459202026 (CS), Lot Number 25DBG283; REF DYNJ906380C: UDI/DI 10198459131042 (EA) 40198459131043 (CS), Lot Number 25FBJ430; REF DYNJ906380C: UDI/DI 10198459131042 (EA) 40198459131043 (CS), Lot Number 25FBC561; REF DYNJ906380C: UDI/DI 10198459131042 (EA) 40198459131043 (CS), Lot Number 25EBK650; REF DYNJ906380C: UDI/DI 10198459131042 (EA) 40198459131043 (CS), Lot Number 25EBE453; REF DYNJ910634: UDI/DI 10198459050398 (EA) 40198459050399 (CS), Lot Number 25CBS527; REF DYNJ910927: UDI/DI 10198459204326 (EA) 40198459204327 (CS), Lot Number 25FMC511; REF DYNJ910927: UDI/DI 10198459204326 (EA) 40198459204327 (CS), Lot Number 25CMJ281.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 21, 2026
Recall initiation date
Dec 16, 2025
Center classification date
Jan 14, 2026

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records