What the record says
- Product
- VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- Reason
- The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
- Classification
- Class I
- Recalling firm
- Ventec Life Systems, Inc.
- Quantity
- 250 breathing circuits
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1070-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Mar 6, 2024
- Recall initiation date
- Jan 9, 2024
- Center classification date
- Feb 23, 2024