11272VN 11272VNK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018); 11272VNU 11272VNUK Flexible Video Urethro C... — FDA record Z-1064-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Apr 1, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
11272VN 11272VNK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018); 11272VNU 11272VNUK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018)
Reason
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Classification
Class II
Recalling firm
Karl Storz Endoscopy
Quantity
All Serial Numbers manufactured/distributed since January 2018
Agency-reported status
Ongoing
Source record identifier
Z-1064-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability11272VN 11272VNK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018) UDI Code: 4048551226186; 11272VNU 11272VNUK Flexible Video Urethro Cystoscope Z18442US-BD (08/2018) UDI Code: 4048551226193; All Serial Numbers manufactured/distributed since January 2018.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY. O.U.S.: Mexico, Lebanon, and South Korea

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
May 18, 2022
Recall initiation date
Apr 1, 2022
Center classification date
May 12, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records