AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001 — FDA record Z-1058-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 10, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001
Reason
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Classification
Class I
Recalling firm
Vyaire Medical
Quantity
66,078 units
Agency-reported status
Ongoing
Source record identifier
Z-1058-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI Case: 50190752114096, Each: 10190752114098; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 6, 2024
Recall initiation date
Jan 10, 2024
Center classification date
Feb 24, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.