Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System — FDA record Z-1049-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 16, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System
Reason
Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
Classification
Class II
Recalling firm
NovaSignal Corp.
Quantity
6
Agency-reported status
Ongoing
Source record identifier
Z-1049-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI/Serial Numbers: 00850681007221/ 21020003, 20080004, 20100004, 21080001, 20120001; 00850681007009/ 19090005

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 8, 2023
Recall initiation date
Dec 16, 2022
Center classification date
Feb 1, 2023