GORE CARDIOFORM Septal Occluder, REF: GSX0030A — FDA record Z-1048-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 14, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
GORE CARDIOFORM Septal Occluder, REF: GSX0030A
Reason
Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder embolization, anoxic brain injury, central venous or arterial perforation or rupture.
Classification
Class II
Recalling firm
W L Gore & Associates, Inc.
Quantity
14
Agency-reported status
Ongoing
Source record identifier
Z-1048-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI/Serial Numbers: 00733132631032/25556388, 25556389, 25556390, 25556391, 25556392, 25556393, 25556394, 25556398, 25556399, 25556400, 25556401, 25556402, 25556403, 25556406

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of NY, CA, FL, CT, TX, AR, IN, MA, MO, SD, VA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 8, 2023
Recall initiation date
Dec 14, 2022
Center classification date
Feb 1, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records