What the record says
- Product
- SMV BodyTrack, System, Tomography, Computed, Emission
- Reason
- GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- Classification
- Class II
- Recalling firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Quantity
- 12 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1046-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI/DI Not applicable: All serial numbers in distribution |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Jan 21, 2026
- Recall initiation date
- Dec 10, 2025
- Center classification date
- Jan 9, 2026
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.