MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008 — FDA record Z-1045-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 22, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008
Reason
The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.
Classification
Class II
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
420 units
Agency-reported status
Ongoing
Source record identifier
Z-1045-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN 10193489120387, Lot Number 810304

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 8, 2023
Recall initiation date
Dec 22, 2022
Center classification date
Feb 1, 2023