BD BBL Sensi Disc Imipenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231644 ... — FDA record Z-1031-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 8, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BD BBL Sensi Disc Imipenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231644 and 231645¿
Reason
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Classification
Class II
Recalling firm
Becton Dickinson & Co.
Quantity
2,363,168 total units
Agency-reported status
Ongoing
Source record identifier
Z-1031-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog No. 231644 UDI-DI 00382902316442 Lots 1335006 2056580 2091918 2210318 2301385 3062101 3094134 3118125 3214950; Catalog No. 231645¿ UDI-DI 30382902316450 Lots 3062101 3094134 3118125 3214950 3261103 1335006 2056580 2081151 2091918 2091925 2147704 2210318 2210343 2277500 2301385

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution nationwide. International distribution worldwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 14, 2024
Recall initiation date
Jan 8, 2024
Center classification date
Feb 8, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records