Recall archive / FDA device enforcement
BD BBL Sensi Disc Ertapenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 232174... — FDA record Z-1030-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jan 8, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- BD BBL Sensi Disc Ertapenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 232174 and 232175¿
- Reason
- There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
- Classification
- Class II
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 2,363,168 total units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1030-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Catalog No. 232174 UDI-DI 00382902321743 Lots 1301766 1341479 2119610 2147329 2242618 2276113 2333166 3152004 3181140; Catalog No. 232175¿ UDI-DI 30382902321751 Lots 1301766 2276113 2333166 2333175 3152004 3181140 1341479 2119610 2119612 2147329 2242618 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Domestic distribution nationwide. International distribution worldwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Feb 14, 2024
- Recall initiation date
- Jan 8, 2024
- Center classification date
- Feb 8, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.