Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB811. Used in C... — FDA record Z-1025-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 12, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB811. Used in Cardiopulmonary bypass.
Reason
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Classification
Class II
Recalling firm
Medtronic Perfusion Systems
Quantity
1,340 units
Agency-reported status
Ongoing
Source record identifier
Z-1025-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel CB811; Serial Number Range: 8111483548 - 8113999999; GTIN Number: 00763000225476 (Lot Numbers: 225313243, 225400271, 225467451, 225467452); GTIN Number: 00763000225483 (Lot Numbers: 13392578, 13394967, 224274665, 224426764, 225332708, 225332709, 225467446, 225467447)

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Belgium, Plurinational State of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Honduras, Hungary, Iceland, Indonesia, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Latvia, Lithuania, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Slovenia, South Africa, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, Turkmenistan, Uganda, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Yemen.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 8, 2023
Recall initiation date
Dec 12, 2022
Center classification date
Jan 27, 2023