Recall archive / FDA device enforcement
Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intratheca... — FDA record Z-1021-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 4, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Model/Catalog Number: 530159
- Reason
- Incorrect catheter connector within the kit resulting in inability to provide a secure connection to the epidural catheter. The incorrect catheter connector is smaller than the required catheter connector and therefore is unable to be used to administer medication.
- Classification
- Class II
- Recalling firm
- B. Braun Medical, Inc.
- Quantity
- 1090 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-1021-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI 04046964669286 (Primary); 04046964669279 (Unit of Use); Lot Number 0061916572 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the states of MO, OK.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Feb 5, 2025
- Recall initiation date
- Dec 4, 2024
- Center classification date
- Jan 24, 2025