Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300 — FDA record Z-1019-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300
Reason
The affected lots show a decline in performance over time, which may lead to false-negative results.
Classification
Class II
Recalling firm
Meridian Bioscience Inc
Quantity
211 units
Agency-reported status
Ongoing
Source record identifier
Z-1019-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00840733102349; Lot Numbers: 410300U195, 410300U196, 410300U197, 410300U201

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution and the OUS country of Italy.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 14, 2026
Recall initiation date
Dec 5, 2025
Center classification date
Jan 5, 2026