Air-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.5 50155 2... — FDA record Z-1019-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 22, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Air-Q3 AirQ3 Standard ILA with Manual Cuff Inflation and GA Size REF Number 0.0 50005 0.5 50055 1.0 50105 1.5 50155 2.0 50205 3.0 50305 4.0 50405
Reason
There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
Classification
Class II
Recalling firm
SunMed Holdings, LLC
Quantity
43,290 (Eaches)
Agency-reported status
Ongoing
Source record identifier
Z-1019-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityALL LOTS Size REF UDI/DI 0.0 50005 Each: 00814954020188 Box: 10814954020185 Case: 20814954020182 0.5 50055 Each: 00814954020195 Box: 10814954020192 Case: 20814954020199 1.0 50105 Each: 00814954020201 Box: 10814954021007 Case: 20814954020205 1.5 50155 Each: 00814954020218 Box: 10814954021014 Case: 20814954020212 2.0 50205 Each: 00814954020225 Box: 10814954021021 Case: 20814954020229 3.0 50305 Each: 00814954020232 Box: 10814954021038 Case: 20814954020236 4.0 50405 Each: 00814954020249 Box: 10814954021045 Case: 20814954020243

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AL, CA, FL, MI, OH, TN, TX, WA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 1, 2023
Recall initiation date
Dec 22, 2022
Center classification date
Jan 26, 2023