BVI Leos Laser and Endoscopy System. Model Number: OME6000U. — FDA record Z-1008-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 5, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Reason
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
Classification
Class II
Recalling firm
Beaver-Visitec International, Inc.
Quantity
5 units
Agency-reported status
Ongoing
Source record identifier
Z-1008-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of FL, MI, OK, TX, WI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 7, 2026
Recall initiation date
Dec 5, 2025
Center classification date
Dec 31, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records