Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MM... — FDA record Z-1005-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MMT-7040QC1, MMT-7040QC2, MMT-7040QC3
Reason
Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 ¿m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot.
Classification
Class II
Recalling firm
Medtronic MiniMed, Inc.
Quantity
114,758
Agency-reported status
Ongoing
Source record identifier
Z-1005-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI/Lot: MMT-7040A/20763000413389/HG72B2C, HG72CQQ, HG73VB7, HG73VC8, HG73VFT, HG73WFR, HG7441E, HG745F9, HG7486X, HG74AJ1, HG74SUR, HG74T9Q, HG74U36, HG78LMS, HG79Y9U; MMT-7040C1/00763000555276, 00763000413392, 20763000413396, 20763000555270, 20763000712475, 00763000712471/F1323, F1923, G0323, G1723, G2023, G2623, G2923, HG6RV7S, HG6TGNU, HG6TJ2Y, HG6TW97, HG6TWVU, HG6UNF2, HG6V8NN, HG6VB8S, HG6VNX3, HG6WNE1, HG6XFPR, HG6XH54, HG6XQCG, HG6XZJX, HG6Y0W1, HG6Y3QN, HG6YEK6, HG6YFAF, HG6YFHL, HG6YJAR, HG6YJBD, HG6YNGD, HG6Z7N5, HG6ZARN, HG7038Y, HG70CW5, HG70DAM, HG70JSU, HG70LRC, HG70QWA, HG71VMZ, HG72DB5, HG72G9C, HG72NV9, HG72R1P, HG72RBD, HG72SE3, HG72TPT, HG72TRV, HG72V3J, HG72ZFP, HG7340A, HG73EJX, HG73RNR, HG745SV, HG747E3, HG7480Q, HG749N6, HG74F93, HG75TSX, HG75XA1, HG76HFH, HG76RYY; MMT-7040C2/20763000555287/HG6RC39, HG6RXB5, HG6T4T7, HG6T8RV, HG6TJ2Y, HG6WKZU, HG6WLKS, HG6XFPR, HG6XH54, HG6XLSD, HG6XNH3, HG6Y5C6, HG6Z1X5, HG713AK, HG73EGX, HG73EJX, HG73FJP, HG73Q24, HG73QPM, HG73RNR, HG7619X, HG764J9, HG76CWC, HG77FMK; MMT-7040C3/20763000555294/HG6XNH3, HG73RNR; MMT-7040D1/00763000413484/HG6THY1, HG6V7VV; MMT-7040D2/00763000413491/HG6TL44; MMT-7040MA/20763000519692/HG74460; MMT-7040QC1/00763000555313, 20763000555317/HG6WT9A, HG6WUGE, HG73PH3; MMT-7040QC2/20763000555324, 763000555320/HG6WT9A; MMT-7040QC3/20763000555331/HG6WT9A

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution including in the states of PR, AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Austria, Belgium, Canary Islands, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Liechtenstein, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 14, 2024
Recall initiation date
Nov 15, 2023
Center classification date
Feb 7, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.