Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2227 — FDA record Z-0985-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 14, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2227
Reason
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Classification
Class II
Recalling firm
Cardinal Health 200, LLC
Quantity
7129 units
Agency-reported status
Ongoing
Source record identifier
Z-0985-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 50885380182592(cs), 20885380182591(ea), Lot Numbers: N210278, N210368, N210423, N210460, N210468, N210469, N210543, N220050, N220099, N220139, N220163

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 25, 2023
Recall initiation date
Dec 14, 2022
Center classification date
Jan 18, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.