Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Co... — FDA record Z-0980-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 27, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Impella catheters - Intravascular micro axial blood pumps that support a patient's circulatory system. (1) Product Code 005042 - Impella 2.5 (2) Product Code 005062 - Impella 5.0 (3) Product Code 005082- Impella LD (7) Product Codes 0550-0008 and 1000100 - Impella 5.5 with SmartAssist (8) Product Code 0048-0032 - Impella CP (9) Product Codes 1000080 and 0048-0045 - Impella CP with SmartAssist ***Updated February 2024*** (10) 004413 - Impella 2.5 Set (12) 0046-0026 - Impella 5.0 Pump Set ROW (13) 0046-0037 - Impella 5.0 Pump Set APAC (15) 004680-AU - 5.0 Pump Set AU (16) 0048-0002 - Impella CP Pump Set, EU (17) 0048-0002-BR - Impella CP Pump Set BR (18) 0048-0004 - Impella CP Pump Set, Canada (19) 0048-0014 - Impella CP Smart Assist Set, EU (20) 0048-0024-JP - Impella CP Smart Assist Set, JP (21) 0048-0044 - Impella CP Smart Assist Set, Canada (22) 0048-0047 - Impella CP Smart Assist Set APAC (23) 005040 - Impella 2.5 IMC Pump Set EU (24) 005048-JP - Impella 2.5 Pump Set, Japan (25) 005060 - Impella 5.0 IMC Pump Set EU (26) 005064 - Impella 5.0 IMC Pump Set Canada (27) 005066-JP - Impella 5.0 Pump Set, Japan (28) 0550-0002 - Impella 5.5 with SmartAssist Set, EU (29) 0550-0004 - Impella 5.5 with SmartAssist Set, CA (30) 1000115 - Impella CP Pump set, APAC (31) 1000211 - Impella 5.5 SmartAssist Set, JP (32) 1000302 - Impella CP with SmartAssist APAC (33) 1000361 - Impella 5.5 Set AU (34) 1000402 - Impella CP Smart Assist Set ***Updated March 15, 2024*** The following products were removed from the list of affected products because the Right-side devices are not in the scope of the warning to avoid left ventricle perforation: (4) Product Code 004334 - Impella RP (5) Product Code 0046-0035 - Impella RP with SmartAssist (6) Product Code 1000323 - RP Flex with SmartAssist (11) 0046-0011 - Impella RP Pump Set, EU (14) 0046-0039 - Impella RP Set APAC
Reason
IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
Classification
Class I
Recalling firm
Abiomed, Inc.
Quantity
91,914 (65,857 US; 26,075 OUS)
Agency-reported status
Ongoing
Source record identifier
Z-0980-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI (1) 00813502011081 (2) 00813502011180 (3) 00813502011227 (7) 00813502011531 and 00813502012828 (8) 00813502011388 (9) 00813502012279 and 00813502011876 All product IFUs include the Update ***Added February 2024*** (10) 813502010947 (12) 813502011821 (13) 813502011937 (15) N/A (16) 4260113630242 (17) 813502011777 (18) 813502011265 (19) 4260113630280 (20) 813502011609 (21) 813502012200 (22) 813502011944 (23) 4260113630136 (24) 813502010046 (25) 4260113630174 (26) 813502011197 (27) 813502010053 (28) 813502011630 (29) 813502010466 (30) 813502010473 (31) 813502012453 (32) 813502012767 (33) 813502012873 (34) 813502012873 ***Removed from scope March 15, 2024*** (4) 00813502011029 (5) 00813502011869 (6) 00813502012811 (11) 4260113630273 (14) 813502011951

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of AU, CA, DE, FR, IN, MX, TW, and VI.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 21, 2024
Recall initiation date
Dec 27, 2023
Center classification date
Feb 9, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.