Artis Q.zen ceiling, Model Number 10848354 — FDA record Z-0979-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 23, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Artis Q.zen ceiling, Model Number 10848354
Reason
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
81 systems
Agency-reported status
Ongoing
Source record identifier
Z-0979-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 4056869010021; Serial Numbers: 111160 111162 111163 111137 111502 111139 111174 111185 111503 111005 111002 111145 111176 111510 111155 111179 111140 111507 111122 111168 111170 111173 111512 111141 111508 111521 111133 111003 111111 111103 111101 111011 111182 111184 111143 111150 111164 111183 111504 111149 111109 111151 111129 111121 111130 111517 111513 111017 111010 111115 111116 111117 111119 111104 111148 111127 111175 111506 111112 111131 111132 111134 111158 111167 111157 111161 111142 111156 111500 111501 111169 111505 111166 111107 111514 111000 111106 111123 111128 111135 Additional Serial Numbers as of 10/4/23: 111125

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 27, 2022
Recall initiation date
Feb 23, 2022
Center classification date
Apr 21, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.