Artis Q zeego, Model Number 10848283 — FDA record Z-0977-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 23, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Artis Q zeego, Model Number 10848283
Reason
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
38 systems
Agency-reported status
Ongoing
Source record identifier
Z-0977-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 4056869010007; Serial Numbers: 117021 117136 117324 117108 117001 117023 117329 117315 117125 117006 117312 117124 117129 117112 117034 117026 117029 117123 117100 117307 117117 117111 117304 117024 117325 117303 117030 117314 117309 117113 117322 117014 117040 117041 117114 117009 Additional Serial Numbers as of 10/4/23: 117127 117310

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 27, 2022
Recall initiation date
Feb 23, 2022
Center classification date
Apr 21, 2022