Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system — FDA record Z-0973-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 6, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Reason
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Classification
Class II
Recalling firm
Philips North America
Quantity
1
Agency-reported status
Ongoing
Source record identifier
Z-0973-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI 00884838009820 Ingenia Serial number 41045 Accessory Serial Number 100018

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 14, 2024
Recall initiation date
Jan 6, 2022
Center classification date
Feb 2, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.