Artis zee floor MN, Model Number 10094142 — FDA record Z-0971-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 23, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Artis zee floor MN, Model Number 10094142
Reason
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
24 systems
Agency-reported status
Ongoing
Source record identifier
Z-0971-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: 4056869010090; Serial Numbers: 140421 140115 140101 140205 140213 140204 140209 140219 140224 140218 140226 140221 140220 140206 140105 140423 140406 140422 140102 140214 140109 140202 140424 140114

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 27, 2022
Recall initiation date
Feb 23, 2022
Center classification date
Apr 21, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.