BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR Catalog Numbers: 546308000 (Size 8); 546310000 (Size 10); 546312000 (Size 1... — FDA record Z-0951-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 19, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR Catalog Numbers: 546308000 (Size 8); 546310000 (Size 10); 546312000 (Size 12); 546314000 (Size 14); 546316000 (Size 16); 546318000 (Size 18); 546320000 (Size 20)
Reason
All lots of BIOSTOP G Bioresorbable Cement Restrictor are being removed as a precautionary measure because tested endotoxin levels were higher than recommended by the current FDA regulatory guidance. In general, endotoxins have a potential to initiate inflammatory responses, ranging from a mild fever to potentially impact or damage to vital organs.
Classification
Class II
Recalling firm
Depuy Ireland UC
Quantity
2199 US; 43515 OUS
Agency-reported status
Ongoing
Source record identifier
Z-0951-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog Number Part Description UDI/DI * 546308000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 8 10603295174547 546310000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 10 10603295174554 546312000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 12 10603295174561 546314000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 14 10603295174578 546316000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 16 10603295174585 546318000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 18 10603295174592 546320000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 20 10603295174608 * UDI-DI on GS1 0603295a0035792 Expiry dates December 2022 thru May 2025 ALL LOTS

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Hungary, Ireland, Israel, Italy, Latvia, Lebanon, Luxemburg, Mexico, Netherlands, New Zealand, Norway, Portugal, Reunion, Romania, Russia, Slovakia, South Africa, Spain, Sweden, Switzerland, and Uruguay.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 18, 2023
Recall initiation date
Dec 19, 2022
Center classification date
Jan 12, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.