Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navi... — FDA record Z-0937-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 17, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355
Reason
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.
Classification
Class II
Recalling firm
SEASPINE ORTHOPEDICS CORPORATION
Quantity
310
Agency-reported status
Ongoing
Source record identifier
Z-0937-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySurgical Technique Guide revisions: D0003926A, D0003926B, D0003926C, D0003926D. REF/UDI-DI/Lot: PC2-400012/10889981260033/MK81527A, MK81527ARR, MK81527ARRR, MK81574AR, MK81574AR1, MK93487A, MK93488ARR, MK93489A, MK93490A, MK93490ARR, MK93491AR, MK93491AR1, MK943488A, MK95882A, MK95882ARR, ML52083B, ML53136B, ML53188B, ML69777B, ML69778B, ML71257B, MM1001498D, MM1001499D, MM1001748D, MM1001749D, MM1001813D, MM1001902D, MM1001903D, MM1002025D, MM1002316D, MM1002328D, MM1002344D; PC2-400013/10889981260040/AL657979C, AL658549C, AL658549CRRR, AL658641C, AL661470C, AL671917D, AL683456F, AL702785F; PC2-400135/10889981260057/BK104899A, BK104899AR, BK104899ARR, BK104899ARRR, BK104899ARRR4, BK106822A, CR49020D, CR49023D, CR49060D, CR51550E, CR51972D, CR56472E; PC2-400140/10889981260064/BK104900A, BK104900AR, BK104900ARRR, BK106825A, BK107806A, BK107806ARRR, BK108366A, BK108366ARRR, CR49822D, CR50158D, CR51973D, CR57547E, CR59411E; PC2-400145/10889981260071/BK104901A, BK104901ARRR, BK106827A, CR49021C, CR49061C, CR51974C; PC2-400155/10889981260088/BK104902A, BK104902ARRR, BK106829A, CR49062C, CR51551C, CR51975C, CR57550E, CR59413E; PC2-400235/10889981260095/BK104903A, BK104903ARRR, BK106831A, BK106831ARRR, BK131159C, CR49620C, CR52371D, CR54729D, CR58013D, CR59414D; PC2-400240/10889981260101/BK104904A, BK104904ARRR, BK106833A, BK131160C, CR49621C, CR52372D, CR59415D; PC2-400245/10889981275495/no lots released; PC2-400345/10889981260118/BK104905A, BK104905ARRR, BK106834A, BK106834ARRR, BK131161C, CR49622C, CR52373D, CR54208D; PC2-400355/10889981260125/BK104906A, BK106835A, CR51764C, CR52853D

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 24, 2025
Recall initiation date
Nov 17, 2025
Center classification date
Dec 17, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.