Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W, VBEAR-3... — FDA record Z-0932-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 28, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W, VBEAR-3-7N, VBEAR-5-7-N, VBEAR-3-7S, VBEAR-5-7S, VBEAR-7-7S, VBEAR-3-7W, VBEAR-5-7W. Intended to be used as an aid to the treatment of certain types of fractures, non-unions or osteotomies.
Reason
Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.
Classification
Class II
Recalling firm
TriMed Inc.
Quantity
270
Agency-reported status
Ongoing
Source record identifier
Z-0932-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI/Lots: VBEAL-3-7N/00842188124907/ME23071, ME23072; VBEAL-5-7N/00842188124914/ME23073; VBEAL-3-7S/00842188124938/ME23031, ME23067, ME23082, ME23097; VBEAL-5-7S/00842188124945/ME23068; VBEAL-7-7S/00842188124952/ME23004, ME23086; VBEAL-3-7W/00842188124969/ME23033; VBEAL-5-7W/00842188124976/ME23036; VBEAR-3-7N/00842188124990/ME23069, ME23084, ME23085; VBEAR-5-7-N/00842188125003/ME23074; VBEAR-3-7S/00842188125027/ME23029, ME23087; VBEAR-5-7S/00842188125034/ME23019; VBEAR-7-7S/00842188125041/ME23022; VBEAR-3-7W/00842188125058/ME23034; VBEAR-5-7W/00842188125065/ME23041, ME23037

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: FL, MI, AL, NY, CA

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 7, 2024
Recall initiation date
Nov 28, 2023
Center classification date
Jan 31, 2024