Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitori... — FDA record Z-0920-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 18, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: Audio Driver version 6.0.9484.1 HP Pro Mini 400 G9 Desktop PC with Realtek HD Audio Driver Officially Released by HP (audio driver version 6.0.9484.1)
Reason
When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.
Classification
Class II
Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Quantity
906 units
Agency-reported status
Ongoing
Source record identifier
Z-0920-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAudio Driver Version: 6.0.9484.1. UDI-DI: 06936415972417. HP Pro Mini 400 G9 Desktop PC. Realtek HD Audio Driver Released by HP. 1. WS host package (mini PC). Part Number 115-050935-00. Quantity 902 2. Exch, Benevision Mini PC Workstation. Part Number E115-050935-00. Quantity 4

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 24, 2025
Recall initiation date
Nov 18, 2025
Center classification date
Dec 12, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records