Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N — FDA record Z-0920-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 1, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N
Reason
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Classification
Class II
Recalling firm
SIN SISTEMA DE IMPLANTE NACIONAL S.A
Quantity
123 (US 98, OUS 25)
Agency-reported status
Ongoing
Source record identifier
Z-0920-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 07899995273437, Lot: X010469757, Expiration: 2028/01

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 22, 2025
Recall initiation date
Sep 1, 2024
Center classification date
Jan 15, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records