Nextra Hammertoe Correction System 4.5 Middle 3.2 Proximal Kit SKU: NX-4532K Orthopedic surgical instrument — FDA record Z-0910-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 10, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Nextra Hammertoe Correction System 4.5 Middle 3.2 Proximal Kit SKU: NX-4532K Orthopedic surgical instrument
Reason
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Classification
Class II
Recalling firm
Nextremity Solutions
Quantity
4 units
Agency-reported status
Terminated
Source record identifier
Z-0910-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Number: 168117318B, 168125017C UDI: 00817701020011

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Distribution US Nationwide, Japan, and the Netherlands

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 20, 2022
Recall initiation date
Feb 10, 2022
Center classification date
Apr 12, 2022
Termination date
Oct 10, 2024