Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD. — FDA record Z-0907-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 13, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.
Reason
Potential weak pouch seal on Panther Fusion GBS assay cartridge, causing potential for invalid or incorrect results leading to a possible delay of treatment.
Classification
Class II
Recalling firm
Hologic, Inc.
Quantity
380
Agency-reported status
Ongoing
Source record identifier
Z-0907-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 15420045510890, Kit Lot 715927, containing Cartridge Lot 624907, Exp. Date 2025-08-15.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

AR, CA, CT, FL, IN, KS, LA, MA, ME, MI, NC, NJ, NV, NY, PR, TX, UT, VT, WA, WV, and Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 22, 2025
Recall initiation date
Dec 13, 2024
Center classification date
Jan 14, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records