Esteem Synergy Stomahesive Skin Barrier, part of Esteem synergy, which is a two-piece ostomy system consisting of an ... — FDA record Z-0903-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 29, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Esteem Synergy Stomahesive Skin Barrier, part of Esteem synergy, which is a two-piece ostomy system consisting of an adhesive skin barrier and an adhesive pouch: ESTEEM SYN FLX WHT16/35MM PRECUT(1X10), REF: 405473; ESTEEM SYN FLX WHT35/35MM PRECUT(1X10), REF: 405479
Reason
Ostomy skin barrier package contains a pre-cut wafer of 28mm instead of 16mm and/or 35mm pre-cut wafers, as stated package labeling. If pre-cut wafers are too big the device cannot be used and if the pre-cut wafer is too small and the device user tries to apply, this could cause abrasion, stomal cut, and bleeding
Classification
Class II
Recalling firm
ConvaTec, Inc
Quantity
240
Agency-reported status
Ongoing
Source record identifier
Z-0903-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF/UDI-DI/Lot: 405473/00768455185399/3J03813, 405479/00768455185450/3J03814

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US distribution to states of: OH, NY, IL

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 22, 2025
Recall initiation date
Nov 29, 2024
Center classification date
Jan 10, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.