Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical chemistry and i... — FDA record Z-0903-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 22, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Liquid Assayed Specific Protein- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2682 Control Level 1
Reason
Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
Classification
Class II
Recalling firm
Randox Laboratories Ltd.
Quantity
146 units US
Agency-reported status
Ongoing
Source record identifier
Z-0903-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN: 05055273204896 Lot Number/Exp. Date: 584LPC 28th Jun 23; 600LPC 28th Nov 23; 611LPC 28th Sep 24;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 11, 2023
Recall initiation date
Nov 22, 2022
Center classification date
Jan 5, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.