Recall archive / FDA device enforcement
ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of... — FDA record Z-0891-2022
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Mar 8, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.
- Reason
- This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).
- Classification
- Class II
- Recalling firm
- Abbott Laboratories
- Quantity
- 402 kits
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0891-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI: (01)00380740003272(17)220531(10)166000 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign) including countries of: AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVI, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, GREECE, IRAQ, ITALY, KAZAKHSTAN, LITHUANIA, LUXEMBOURG, MACEDONIA, MOROCCO, PANAMA, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, and TURKEY
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 13, 2022
- Recall initiation date
- Mar 8, 2022
- Center classification date
- Apr 7, 2022