Recall archive / FDA device enforcement
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, ... — FDA record Z-0885-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Dec 19, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-32, 0998-UC-0800-33, 0998-UC-0800-34, 0998-UC-0800-35, 0998-UC-0800-36, 0998-UC-0800-45, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55, 0998-UC-0800-65.
- Reason
- The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.
- Classification
- Class I
- Recalling firm
- Datascope Corp.
- Quantity
- 11,906 Total
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0885-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-31, 10607567109053; 0998-00-0800-32, 10607567111117; 0998-00-0800-33, 10607567109053; 0998-00-0800-34, 10607567111940; 0998-00-0800-35, 10607567109107; 0998-00-0800-36, 10607567114187; 0998-00-0800-45, 10607567108421; 0998-00-0800-52, 10607567108438; 0998-00-0800-53, 10607567108391; 0998-00-0800-55, 10607567108414; 0998-00-0800-65, 10607567113432; 0998-UC-0800-31, 10607567109053; 0998-UC-0800-32, 10607567111117; 0998-UC-0800-33, 10607567109008; 0998-UC-0800-34, 10607567111940; 0998-UC-0800-35, 10607567109107; 0998-UC-0800-36, 10607567114187; 0998-UC-0800-45, 10607567108421; 0998-UC-0800-52, 10607567108438; 0998-UC-0800-53, 10607567108391; 0998-UC-0800-55, 10607567108414; 0998-UC-0800-65, 10607567113432. |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide Distribution - US Nationwide and the countries of Albania, Angola, Argentina, Australia, Austria, AZERBAIJAN, BAHRAIN, BANGLADESH, Barbados, Belarus, Belgium, BOSNIA and HERZEGOVINA, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d'Ivoire, Croatia, CZECH REPUBLIC, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, TRINIDAD AND TOBAGO, Tunisia, Turkey, Turkmenistan, Ukraine, UNITED ARAB EMIRATES, United Kingdom, Vietnam, Yemen, and Zambia.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Jan 25, 2023
- Recall initiation date
- Dec 19, 2022
- Center classification date
- Jan 19, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.