Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12) — FDA record Z-0861-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 8, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Reason
Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
Classification
Class II
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
108 systems (14 US, 1 Canada, 93 ROW)
Agency-reported status
Ongoing
Source record identifier
Z-0861-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability(1) System Code: 722079; (1) UDI/DI: 00884838085268; (1) Serial Number: 1005, 866, 1361, 952, 1241, 1169, 756, 703351, 1135, 1122, 824, 748, 630, 638, 703385, 812, 703346, 1303, 1313, 893, 768, 929, 703638, 1357, 703810, 934, 1202, 703461, 703706, 1172, 703959, 703845, 703705; (2) System Code: 722224; (2) UDI/DI: 00884838099258; (2) Serial Number: 539, 395, 278, 134, 730, 252, 370, 36, 572, 645, 272, 495, 781, 62, 1388, 352, 1655, 690, 555, 895, 2165, 861, 60, 20, 648, 2231, 1778, 1431, 778, 477, 917, 1718, 388, 2042, 150, 764, 2036, 1421, 157, 879, 212, 144, 1463, 347, 575, 2187, 355, 2576, 1332, 424, 387, 863, 211, 474, 712, 549, 2309, 799, 2283, 639, 1700, 2071, 537, 1589, 2208, 2057, 771, 772, 1216, 1667;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 7, 2024
Recall initiation date
Dec 8, 2023
Center classification date
Jan 30, 2024