MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762,... — FDA record Z-0855-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Jan 27, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 620G (OUS Only): MMT-1750; MiniMed 640G (OUS Only): MMT-1711, MMT-1712, MMT-1751, MMT-1752; MiniMed 720G (OUS Only): MMT-1809, MMT-1810, MMT-1859, MMT-1860; MiniMed 740G (OUS Only): MMT-1811, MMT-1812, MMT-1861, MMT-1862; MiniMed 780G (OUS Only): MMT-1884, MMT-1885, MMT-1886, MMT-1895, MMT-1896 The MiniMed" 630G System with SmartGuard is intended for continuous delivery of basal insulin (at user selected rates) and administration of insulin boluses (in user selectable amounts) for the management of Type 1 diabetes mellitus in persons sixteen years of age and older.
Reason
New or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.), which must be set up and saved on pumps prior to use. Patients who do not realize this will receive no basal insulin, which can lead to insulin under-delivery. Insulin under-delivery can cause severe hyperglycemia, which may lead to diabetic ketoacidosis. Current device labeling does not clearly identify that these insulin pumps are capable of functioning with the default basal setting of 0.0 units/hr for all 24 hours and that no error message will appear if the default basal setting remains active. Furthermore, the labeling for the affected devices does not highlight that saving the basal settings requires multiple steps and failure to complete all steps will not save the new basal settings, nor that the steps required to set and save a basal profile and other parameters which are used as part of the bolus calculator function are different.
Classification
Class II
Recalling firm
Medtronic MiniMed
Quantity
175,801
Agency-reported status
Ongoing
Source record identifier
Z-0855-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityInsulin pumps distributed in the last 6 months. All devices (US) serial numbers ranging from NG2056884H TO NG2891033H and (OUS) serial numbers ranging from NG1921788H TO NG2883091

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: OH, UT, IL, NY, FL, LA, NE, NV, CA, VA, TX, DE, KS, MS, MI, AL, OK, AZ, SC, IN, AR, PA, WV, MA, RI, NJ, CO, TN, KY, GA, WI, MN, NM, MD, IA, NC, WY, MO, WA, OR, CT, ND, ME, VT, ID, SD, AK, MT, NH, DC, HI, PR, VI OUS: Australia, Japan, Korea, New Zealand, Taiwan, Viet Nam

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 6, 2022
Recall initiation date
Jan 27, 2022
Center classification date
Mar 31, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.