Recall archive / FDA device enforcement
Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system — FDA record Z-0841-2022
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Jan 31, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system
- Reason
- Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement. If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.
- Classification
- Class II
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 70 Systems
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0841-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Serial Numbers: 20113 23071 20118 22523 23021 22169 23164 22388 22119 22844 21209 23699 21114 22190 22146 22126 22128 22836 22163 22727 22135 22347 21174 22676 22029 22895 22917 22920 22058 22847 22852 22655 22708 22275 22254 22928 22857 22208 22071 23130 22092 22175 21194 23602 23018 23019 22076 22630 21108 22919 22825 22054 22177 20105 22998 20115 22104 22966 23625 23627 22748 22352 22353 22354 22618 21004 21005 22136 23059 22914 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Apr 6, 2022
- Recall initiation date
- Jan 31, 2022
- Center classification date
- Mar 30, 2022
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.