Nimbus PainPro Ambulatory Infusion Pumps with software version v4, Model Number IT1072, sold as kit IT1075-DUKE-HSP, ... — FDA record Z-0832-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 19, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Nimbus PainPro Ambulatory Infusion Pumps with software version v4, Model Number IT1072, sold as kit IT1075-DUKE-HSP, IT1076-DUKE-HSP and IT1077-DUKE-HSP
Reason
An unreleased Version 4 Protocol Library was installed on Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients.
Classification
Class II
Recalling firm
InfuTronix LLC
Quantity
74 units
Agency-reported status
Ongoing
Source record identifier
Z-0832-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability(UDI): (01)00817170020086; Lot: 2009231440; Serial Numbers: 807483, 807484, 807485, 807486, 807487, 807489, 807490, 807491, 807492, 807493, 807494, 807495, 807496, 807497, 807498, 807499, 807500, 807502, 807651, 807652, 807653, 807507, 807508, 807509, 807510, 807511, 807512, 807513, 807514, 807515, 807516, 807517, 807518, 807519, 807520, 807521, 807522, 807523, 807524, 807525, 807526, 807527, 807528, 807529, 807530, 807531, 807532, 807533, 807534, 807535, 807536, 807537, 807538, 807539, 807540, 807541, 807542, 807543, 807544, 807545, 807546, 807547, 807548, 807549, 807550, 807551, 807552, 807553, 807554, 807654, 807655, 807656, 807657, 807658

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of NC and OH.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Apr 6, 2022
Recall initiation date
Oct 19, 2021
Center classification date
Mar 29, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records