What the record says

Product
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L; b) OB/GYN DAVINCI PACK A RG BASIN, kit number HGOB98G; c) DAVINCI GYN PACK, kit number HHDV17; d) KIT, GYNE ROBOTIC, kit number MMGR44D; e) KIT, MAJOR LITHOTOMY, kit number MMJL16E; f) GYN PACK, kit number SMGY46; g) Lithotomy Pack, kit number THLH71C; h) ASC HYSTEROSCOPY PACK (PS 020634), kit number UIHY16AA; i) GYN LAPAROSCOPY HPES, kit number WELP67C; j) GYN LAPAROSCOPY HPES, kit number WELP67C-01; k) GYN LAPAROSCOPY HPES, kit number WELP67C-02
Reason
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Classification
Class II
Recalling firm
American Contract Systems, Inc.
Quantity
1020 units
Agency-reported status
Ongoing
Source record identifier
Z-0828-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitya) AGLP52L, UDI/DI 00191072144119, Lot Numbers: 933231, exp. 3/8/2024; 942231, exp. 2/27/2024; 947231, exp. 2/22/2024; 966231, exp. 2/3/2024; 980231, exp. 1/20/2024; 681221, exp. 11/15/2023; 707221, exp. 10/20/2023; b) HGOB98G, UDI/DI none, Lot Numbers: 667221, exp. 11/29/2023; 977231, exp. 1/23/2024; 952231, exp. 1/31/2024; 896231, exp. 1/31/2024; c) HHDV17, UDI/DI 00191072170125, Lot Numbers: 766221, exp. 8/22/2023; 731221, exp. 9/26/2023; 682221, exp. 11/14/2023; 638221, exp. 12/28/2023; 946231, exp. 2/23/2024; d) MMGR44D, UDI/DI none, Lot Numbers: 709221, exp. 10/18/2023; 650221, exp. 12/16/2023; 891231, exp. 4/19/2024; e) MMJL16E, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 920231, exp. 3/21/2024; f) SMGY46, UDI/DI 00191072188458, Lot Numbers: 961231, exp. 2/8/2024; g) THLH71C, UDI/DI 00191072155108, Lot Numbers: 756221, exp. 9/1/2023; 708221, exp. 10/19/2023; h) UIHY16AA, UDI/DI none, Lot Numbers: 931231, exp. 3/10/2024; 896231, exp. 4/14/2024; i) WELP67C, UDI/DI 00191072173676, Lot Numbers: 751221, exp. 9/6/2023; 681221, exp. 9/13/2023; j) WELP67C-01, UDI/DI 00191072173676, Lot Numbers: 720221, exp. 9/21/2023; k) WELP67C-02, UDI/DI 00191072173676, Lot Numbers: 643221, exp. 9/13/2023

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Feb 7, 2024
Recall initiation date
Nov 29, 2023
Center classification date
Jan 26, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records