Achieva 3.0T, Model No. 781177 — FDA record Z-0809-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 29, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Achieva 3.0T, Model No. 781177
Reason
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Classification
Class III
Recalling firm
Philips North America
Quantity
7 US; 16 ROW total
Agency-reported status
Ongoing
Source record identifier
Z-0809-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel No. 781177 UDI-DI N/A Serial Number 17043

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 31, 2024
Recall initiation date
Dec 29, 2023
Center classification date
Jan 25, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.