BioPlex 2200 REF 12000650 SYPH T PACK Syphilis Total & RPR Reagent Pack — FDA record Z-0806-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Feb 8, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BioPlex 2200 REF 12000650 SYPH T PACK Syphilis Total & RPR Reagent Pack
Reason
Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent manufacturing challenges to produce RPR reagents that consistently meet stability specifications
Classification
Class II
Recalling firm
Bio-Rad Laboratories, Inc.
Quantity
9,426 packs (U.S.)
Agency-reported status
Ongoing
Source record identifier
Z-0806-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog Number: 12000650 UDI GTIN Code: 03610520653012 Lot Numbers: 301336

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of France, Germany, Italy, Spain, Saudi Arabia, and United Arab Emirates.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Mar 23, 2022
Recall initiation date
Feb 8, 2022
Center classification date
Mar 16, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records