EliA GBM Wells, Material Number 14551401, for Phadia 250, 2500, and 5000. In vitro diagnostic — FDA record Z-0783-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date May 24, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
EliA GBM Wells, Material Number 14551401, for Phadia 250, 2500, and 5000. In vitro diagnostic
Reason
Complaints that specific samples produced false positive EliA GBM results. A positive signal was present when these samples were tested for coating-solution reactivity using EliA wells without antigen.
Classification
Class II
Recalling firm
Phadia Ab
Quantity
124,928 kits
Agency-reported status
Ongoing
Source record identifier
Z-0783-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 07333066010670; All Lots.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 31, 2024
Recall initiation date
May 24, 2023
Center classification date
Jan 24, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.