Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product Code 35700BAX2 — FDA record Z-0776-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 19, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product Code 35700BAX2
Reason
Devices were identified as released after repair without full testing being performed, which includes flow testing.
Classification
Class II
Recalling firm
Baxter Healthcare Corporation
Quantity
1 unit
Agency-reported status
Ongoing
Source record identifier
Z-0776-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN 00085412498683; Serial Number: 2153638

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of AL, IN.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 8, 2025
Recall initiation date
Nov 19, 2024
Center classification date
Dec 27, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records