Guardian iOS app (MMT-8200) and Guardian Android app (MMT-8201), part of the Guardian 4 system — FDA record Z-0776-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 21, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Guardian iOS app (MMT-8200) and Guardian Android app (MMT-8201), part of the Guardian 4 system
Reason
An app, part of a continuous glucose monitoring system, for use with smartphone devices may automatically log out from CareLink, then the app is not able to upload data. When logged out, linked care partners will not receive SMS notifications (could result in hypoglycemia or hyperglycemia), and sensor glucose values will also not be sent to the InPen app.
Classification
Class II
Recalling firm
Medtronic MiniMed
Quantity
3235
Agency-reported status
Ongoing
Source record identifier
Z-0776-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityApp Version 1.3.0

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

OUS: Austria, Belgium, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Luxembourg, Moldova, Republic of, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Jan 11, 2023
Recall initiation date
Nov 21, 2022
Center classification date
Dec 31, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records