EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002 — FDA record Z-0775-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Dec 29, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002
Reason
GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.
Classification
Class I
Recalling firm
Datex-Ohmeda, Inc.
Quantity
12 units
Agency-reported status
Ongoing
Source record identifier
Z-0775-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI ********, Serial Numbers: CBCS01269, CBCT02328, CBCT03050, CBCT02306, CBCT02301, CBCT02303, CBCT02302, CBCT02088, CBCT02309, CBCT02304, CBCU00717, CBCU00559

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Feb 7, 2024
Recall initiation date
Dec 29, 2023
Center classification date
Jan 30, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records